Location: Hammersmith campus, Victor Phillip Dahdaleh building, 2nd floor, lab VPD 206A.

This sorter is operated with 70um, 85um, or 100um nozzle. The model also fits 130um nozzle (suitable for very large and/or fragile cells), but the facility does not currently offer this option.

Can sort up to 4-way into tubes, and also sorts into 96-well plates.

Filter configuration:

Laser Mirror Filter Common fluorochromes, dyes, fluorescent proteins
355nm UV 450LP 515/30 BUV496, LD Aqua (DAPI, Hoechst 33342, Zombie UV, LD Blue)
- 379/28 BUV395
405nm Violet 750LP 780/60 BV750, BV785
690LP 710/50 BV711
630LP 660/20 BV650
595LP 610/20 BV605
505LP 525/50 Zombie Aqua, BV510, BV480
- 450/50 DAPI, BV421, Pacific Blue, eF450, BFP, LD Violet
488nm Blue 685LP 710/50 PerCP-Cy5.5, PerCP-eF710, PE-Cy5.5, BB700, RB705
505LP 530/30 FITC, AF488, GFP, BB515
561nm Yellow 735LP 780/60 PE-Cy7
685LP 710/50 PerCP-Cy5.5, PerCP-eF710, BB700
630LP 670/25 PE-Cy5, 7-AAD
600LP 610/20 PE-TexasRed, PE-CF594, mCherry
- 582/15 PE, tdTomato
640nm Red 750LP 780/60 APCFire750, APCFire810, AF750, APC-Cy7, APC-eF780
690LP 730/45 AF680, AF700, DRAQ5, DRAQ7
- 670/30 APC, AF647, TO-PRO-3

Some spare filters are available and, depending on the full panel, it might be possible to use BUV563, BUV615, BUV661, BUV737, BUV805, etc. Please enquire with staff.

Compensation can be calculated automatically or manually based on single-stained controls and can also be edited manually.

Software: BD FACSDiva

Manufacturer website: BD FACSAria Fusion

Department of Immunology and Inflammation

General enquiries


Hammersmith Flow Cytometry Facility
Dr Victor Phillip Dahdaleh (VPD) building
Level 2, Lab 206 and Office 205

Hammersmith Hospital
Du Cane Road
London W12 0NN

Facility Managers
Joana De Teixeira Carrelha
Email: j.de-teixeira-carrelha@imperial.ac.uk

Fredrik Wållberg
Email: f.wallberg@imperial.ac.uk